EZ Match User Manual
3-7
Tube/Cap Inspection
Timeout (Multi-Operation)
Note: Multi-Operation Timeout applies only to the specified operation (1 through 3).
Database Range (Multi-Operation)
When Multi-Operation is active, the Configuration Datase cannot be disabled. The database
range fields specify the start and end index range to be used for each operation.
Minimum
Maximum
Definition: Timeout indicates the amount of time required before the read cycle will
switch to the next operation. Timeout is configurable in 1 ms increments.
Serial Cmd: <K260,index,operation,frame count,timeout,database min,database max>
Options: 1 to 65535 (in 1 ms increments)
Definition:
Specifies the starting IP Database index to be used for processing
during the
specified operation.
Serial Cmd: <K260,index,operation,frame count,timeout,database min,database max>
Options: 1 to 10
Definition: Specifies the ending IP Database index to be used for processing during the
specified operation. If this value is set to less than the minimum value, the
minimum value will be used for both max and min.
Serial Cmd: <K260,index,operation,frame count,timeout,database min,database max>
Options: 1 to 10