Chapter 15 Diagnostics
MS-880 Industrial Long Range Scanner User’s Manual 15-11
15–Diagnostics
Present Ambient Temperature (deg. C)
(This command is READ ONLY.)
Note: When viewing the Present Operating Temperature, allow at least 20 min-
utes after startup for the scanner’s inside and outside temperatures to stabilize.
Definition: Close estimation of external temperature.
A thermal sensor on the main PCB measures the internal temperature
of the scanner. From this a case-differential constant is subtracted to
yield an estimated external temperature.
Usage: Indicates temperature of the environment immediately outside of the
scanner. Useful for setting Over Temperature Threshold and Under
Temperature Threshold.
Serial Cmd: Send: <K408?>
Returns <K408 degrees Celsius>.