I/O Parameters
Dimension 1
Dimension 2
Dimension Tolerance
Definition: Checks symbol dimension for a specified number of pixels. In the case of
rectangular symbols, checks the longer side.
Serial Cmd: <K718,finder pattern status,symbol size mode,symbol size 1,symbol size
2,symbol size tolerance,dimension mode,dimension 1,dimension
2,dimension tolerance,orientation mode,orientation value>
Default: 0
Options: 0 to 640 pixels
Definition: Checks symbol dimension for a specified number of pixels. In the case of
rectangular symbols, checks the shorter side.
Serial Cmd: <K718,finder pattern status,symbol size mode,symbol size 1,symbol size
2,symbol size tolerance,dimension mode,dimension 1,dimension
2,dimension tolerance,orientation mode,orientation value>
Default: 0
Options: 0 to 480 pixels
Definition: Sets the allowable deviation, in percentage, for symbol sizes specified in
Dimension 1 and Dimension 2.
Serial Cmd: <K718,finder pattern status,symbol size mode,symbol size 1,symbol size
2,symbol size tolerance,dimension mode,dimension 1,dimension
2,dimension tolerance,orientation mode,orientation value>
Default: 10
Options: 0 to 100