MS-4 Imager User’s Manual 4-17
Calibration
Additional Notes about Calibration
The following conditions apply to the Imager’s calibration process. Some of these items
are noted at various points throughout this section, or in other sections of the Imager’s
documentation.
1. The Mirrored Image parameter <K514> is not part of the calibration process.
2. Background Color <K451> is calibrated.
3. If Window of Interest (WOI) Framing is enabled, the WOI will be set to full frame
when calibration begins. If WOI Framing is disabled, the current WOI configuration
will be used for the Search Pass.
4. IP Mode <K527> will not be altered during calibration.
5.
Threshold Mode
<K512>
will be set to
Adaptive
during calibration. At the end of a
successful calibration,
Threshold Mode
will remain set to
Adaptive
regardless of its
previous setting. Otherwise, the value will revert to its original state.
6. If Symbol Type is enabled for calibration (Autodiscriminate), then Interleaved 2 of 5
Range Mode Status (Interleaved 2 of 5)
<K472>
will be enabled. This allows variable-length
Interleaved 2 of 5 symbols to be decoded. If an Interleaved 2 of 5 symbol is decoded
during calibration, then symbol length # 1 will be set to the decoded symbol length at
the end of calibration. Otherwise, the symbol lengths will be restored to their original
configurations.
7. The symbol size, dimension, and orientation parameters of the No Symbol Message
and Bad Symbol Message, and 2D Symbology Qualification command <K718>
will be updated if calibration is completed successfully.
8. All symbol types that were enabled before calibration will still be enabled after calibration.
For example, if Data Matrix ECC 200 was enabled before calibration, and the calibration
routine was performed on a Code 128 symbol, then after calibration is completed both
Data Matrix ECC 200 and Code 128 will be enabled.
9. The calibration process does not modify the global Composite status <K453>. The
global Composite status must be configured properly before calibration.
10. If the user requires calibration of a stacked symbology <K482>, <K483>, or <K484>,
that symbology must be enabled and configured appropriately before calibration.
11. UPC Supplementals will be disabled during calibration.
12. The Search Pass process will use the configurable Window of Interest for image cap-
tures. However, when the search process is completed, the WOI will be reduced to
include only the symbol of interest and some additional boundary area.